![GSi TYMPSTAR 1 Service Manual Download Page 27](http://html1.mh-extra.com/html/gsi/tympstar-1/tympstar-1_service-manual_2275118027.webp)
Specifications
Calibration Stability
All GSI TympStar Version 1 specifications are met over the range of specified power
line, temperature and humidity variations.
Power Line:
Voltage Variation:
±10%
Frequency Variation:
± 5%
Power line short term variation which affects the performance of the instru-
ment will turn off all probe and stimulus signals.
Power Rating:
120 Watts maximum
Line Voltage Range:
100 VAC to 240 VAC
Power Line Frequency Range:
50 - 60 Hz
Temperature Operating Range:
+15°C to +35°C
Relative Humidity Operating Limit: 90%
Guaranteed Operating Elevation:
6000 Ft. (1800m)
Connectors
STIMULUS: External Stimulus Input
(Phone Jack) Peak Voltage: 3 VAC
Input Impedance: 15,000 Ohms
PRESENT: External Present Control Input that turns the stimulus signal ON
and OFF (Phone Jack).
Voltage Range: STIM OFF: +5.0 VDC
STIM ON: 0.0 v
Input Impedance: 11,000 Ohms
Contra PHONE: Output Voltage: 7 VAC
Output Impedance: 2.5 Ohms
Standard 9-pin RS232C serial port for interfacing with outside computer.
RXD
CTS
TXD
RTS
Standard VGAport for external monitor.
GSI TympStar Version 1 and Version 2 Service Manual
2 - 7
Summary of Contents for TYMPSTAR 1
Page 1: ...Part Number 2000 0110RevB...
Page 20: ...Chapter 1 1 8 Grason Stadler...
Page 72: ...Chapter 3 Grason Stadler 3 32...
Page 168: ...Chapter 5 Grason Stadler 5 30...
Page 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Page 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Page 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Page 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Page 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Page 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Page 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Page 178: ...Chapter 6 Grason Stadler 6 10...
Page 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Page 180: ...Chapter 6 Grason Stadler 6 12...