![GSi TYMPSTAR 1 Service Manual Download Page 24](http://html1.mh-extra.com/html/gsi/tympstar-1/tympstar-1_service-manual_2275118024.webp)
Chapter 2
Intensity levels are reduced as a function of volume at a rate of 1 dB SPL for each .1
ml. Intensity reduction begins at 1.2 ml.
Table 3: Upper limit of HL range in Reflex Threshold Mode
110
105
Lower limit of HL range for all stimulus signals (35 dB HL).
Hearing Level Increment:
5.0 dB
Hearing Level Increment Accuracy:
±0.5 dB
Hearing Level Control Linearity:
±1.0 dB