![GSi TYMPSTAR 1 Service Manual Download Page 211](http://html1.mh-extra.com/html/gsi/tympstar-1/tympstar-1_service-manual_2275118211.webp)
Troubleshooting
ALERT PUMP LIMIT REACHED - RESELECT SOFTKEY
The pump ran out of stroke trying to reach or maintain a specified pressure because
of a leak in the pressure system. The test can be restarted by reselecting the appro-
priate softkey.
ALERT PUMP OVER PRESSURIZED - RESELECT SOFTKEY
The pump pressurized beyond the acceptable pressure range. This condition is de-
tected by software or an over-pressure valve tripping. The test may be restarted by
reselecting the appropriate softkey.
ALERT PUMP LIMIT REACHED
Enter the Pressure Cal portion of the CAL Mode. Select the -600/400 softkey and
perform the Leak Rate test. If leak is excessive, enter the Diagnostic section of the
Pressure Cal Mode.
ERROR INVALID SELECTION - PROBE TONE NOT SELECTED
The Present Bar was pressed to turn on the probe tone without a probe tone fre-
quency being selected during the probe tone Cal Mode.
ALERT ALTITUDE LIMIT REACHED
An altitude was selected beyond the valid range.
ALERT HIGH INTENSITYSELECTED
A stimulus intensity level above 100 dB HL has been selected.
ALERT VALVE NOT TRIPPED -ADJUST VALVE LIMIT SETTING
Enter the Pressure Cal portion of the CALMode, select the Verify Cal softkey, then
the Safety Valve softkey and perform the calibration.
NOTE CANNOT START TEST - PERFORM ALTITUDE CAL - CON-
TACT SERVICE
Perform an altitude calibration.
ALERT PROBE MUST BE IN A CAVITY BEFORE PRESSING DATA
TRANSFER
The probe was detected out of the cavity while attempting to calibrate the compli-
ance.
GSI TympStar Version 1 and Version 2 Service Manual
8 - 21
Summary of Contents for TYMPSTAR 1
Page 1: ...Part Number 2000 0110RevB...
Page 20: ...Chapter 1 1 8 Grason Stadler...
Page 72: ...Chapter 3 Grason Stadler 3 32...
Page 168: ...Chapter 5 Grason Stadler 5 30...
Page 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Page 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Page 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Page 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Page 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Page 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Page 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Page 178: ...Chapter 6 Grason Stadler 6 10...
Page 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Page 180: ...Chapter 6 Grason Stadler 6 12...