Description
DPRAM - Dual Port RAM memory test failure
Subcodes: None.
Likely Causes: Most likely hardware failure with DPRAM on
Digital board or any address/data lines that run to the
DPRAM.
Invalid keycode detected – keycode out of range.
Subcodes: keycode detected.
Likely Causes: This is one of the internal checks for the
keyboard handler. This indicates an internal software failure.
Invalid keycode handler function detected by menu handler.
Subcodes: keycode that is currently being processed.
Likely Causes: This is one of the internal checks for the
keyboard handler. This indicates an internal software failure.
The key queue has filled and additional keys are detected that
cannot be stored by the software.
Subcodes: keycode of the last key that attempted entry into
the queue.
Likely Causes: The key queue is very large and keys cannot
be pressed fast enough to fill it. This is most likely a
hardware problem with keys. It could also mean that the CP
software (PC104) has stopped processing keys, but continues
to queue them up until the buffer fills (this is less likely).
SP does not respond to communication request by CP
Subcodes: none
Likely Cause: When the system is first booted, the DPRAM
is enabled and the SP is taken out of reset (i.e. initialized) and
must respond to the CP’s communication request within 30
seconds. If the SP does not respond, there could be an
internal SP problem or the hardware path between the SP and
CP could be faulty.
Timeout on wait for SP to complete the last set of commands.
Subcodes: command opcode
Likely Cause: There are numerous commands that travel
between the CP and SP; if the SP is not responding for any
reason this error code will likely occur. It is most likely a
problem with the SP itself (for example, if the SP gets reset
by high-current power spike) or with a faulty hardware path
between the SP and CP.
Grason-Stadler
Chapter 8
Error Codes
Error codes are alphanumeric and are generated by the Control Processor (CP) on
the PC 104 board and the Signal Processor (SP) on the Digital board. The Control
Processor is the main host computer and controls the overall operation of the instru-
ment. The Signal Processor controls the signal routing and measuremant process.
CP Error codes
Code
Name
1000
ABORT_DPRAM_FAIL
1100
ABORT_KEY_INVALID
1101
ABORT_KEY_PROC_BAD
1102
ABORT_KEY_TOO_MANY_KEYS
1200
ABORT_SP_FAIL
1201
ABORT_SP_TIMEOUT_ON_CMD
8 - 6
Summary of Contents for TYMPSTAR 1
Page 1: ...Part Number 2000 0110RevB...
Page 20: ...Chapter 1 1 8 Grason Stadler...
Page 72: ...Chapter 3 Grason Stadler 3 32...
Page 168: ...Chapter 5 Grason Stadler 5 30...
Page 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Page 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Page 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Page 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Page 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Page 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Page 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Page 178: ...Chapter 6 Grason Stadler 6 10...
Page 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Page 180: ...Chapter 6 Grason Stadler 6 12...