ABORT_PROBE_GAIN_TIMEOUT
Timeout on wait for SP to set the probe tone gain to the
specified level.
Subcodes:
0 = SP does not complete 75 dB AGC (3 seconds)
1 = SP does not start sending compliance data after
completing AGC (15 seconds)
Likely Causes: SP failure is the most likely cause or
problems in the hardware path between SP and CP (including
DPRAM). An erroneous SP reset could also account for the
problem.
ABORT_PROBE_GAIN_INVALID_DB_SPL
Invalid SPL level parameter received in attempting to set
probe gain.
Subcodes: received SPL level parameter
Likely Causes: This limit check should never be exceeded
(the only allowed values are 75dbSPL and 85dbSPL) so the
most likely cause for this failure would be an internal
software failure such as a memory/stack corruption.
ABORT_REMOTE_TXTEST_FAIL
Unsuccessful call to the communication task layer for
transmission of test data.
Subcodes: none
Likely Causes: The communications task layer has indicated
that it could not process the request in the current state.
Protection at the high level should prevent this from
happening (for example, you cannot transfer data remotely
while printing) but if this low level should detect a problem
then this failsafe code can occur. This indicates an internal
software failure.
ABORT_REMOTE_TXTEST_FAIL2
Unsuccessful call to the communication task layer for
transmission of test data.
Subcodes: none
Likely Causes: The communications task layer has indicated
that it could not process the request in the current state.
Protection at the high level should prevent this from
happening (for example, you cannot transfer data remotely
while printing) but if this low level should detect a problem
then this failsafe code can occur. This indicates an internal
software failure.
ABORT_REMOTE_TXTEST_FAIL3
Unsuccessful call to the communication task layer for
transmission of test data.
Subcodes: none
Likely Causes: The communications task layer has indicated
that it could not process the request in the current state.
Protection at the high level should prevent this from
happening (for example, you cannot transfer data remotely
while printing) but if this low level should detect a problem
then this failsafe code can occur. This indicates an internal
software failure.
ABORT_REMOTE_TXDONE_FAIL
Unsuccessful call to the communications task layer when
transmission is completed.
Subcodes: none
Likely Causes: The communications task layer has indicated
that it could not process the request in the current state.
Protection at the high level should prevent this from
happening (for example, you cannot transfer data remotely
while printing) but if this low level should detect a problem
then this failsafe code can occur. This indicates an internal
software failure.
Grason-Stadler
Chapter 8
CP Error codes
1700
1701
1900
1901
1902
1903
8 - 8
Summary of Contents for TYMPSTAR 1
Page 1: ...Part Number 2000 0110RevB...
Page 20: ...Chapter 1 1 8 Grason Stadler...
Page 72: ...Chapter 3 Grason Stadler 3 32...
Page 168: ...Chapter 5 Grason Stadler 5 30...
Page 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Page 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Page 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Page 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Page 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Page 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Page 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Page 178: ...Chapter 6 Grason Stadler 6 10...
Page 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Page 180: ...Chapter 6 Grason Stadler 6 12...