Calibration
Step 2
Follow the instructions on the screen to
measure the ambient pressure calibra-
tion using an electronic manometer. The
manometer and LCD meter should agree
within ± 10 daPa.
Adjusting the probe
Step 3
If it is necessary to adjust the probe’s
ambient pressure potentiometer:
• Loosen the cable strain relief by hand.
• Remove the two metal, and two nylon screws.
CAUTION
These screws must be reinstalled in the same positions later.
• Carefully remove the probe cover.
CAUTION
Do not disconnect the shield wire.
• Adjust the ambient potentiometer
as shown to achieve an acceptable
reading on the LCD meter.
Step 4
Proceed to the set -600/400 calibration
function or reinstall the cover and press
RETURN to return to the Main Calibra-
tion Menu.
GSI TympStar Version 1 and Version 2 Service Manual
4 - 43
Summary of Contents for TYMPSTAR 1
Page 1: ...Part Number 2000 0110RevB...
Page 20: ...Chapter 1 1 8 Grason Stadler...
Page 72: ...Chapter 3 Grason Stadler 3 32...
Page 168: ...Chapter 5 Grason Stadler 5 30...
Page 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Page 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Page 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Page 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Page 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Page 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Page 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Page 178: ...Chapter 6 Grason Stadler 6 10...
Page 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Page 180: ...Chapter 6 Grason Stadler 6 12...