Chapter 4
The probe tone can be toggled between STANDARD and MULTI-FREQ by press-
ing the STANDARD/MF softkey. This example shows a standard probe tone of
1000 Hz selected.
Selected input filters are displayed in the FILTER list in the center of the LCD, and
can be changed by pressing the FILTER MODE UP and DOWN NO CAPS. This
example shows a filter mode of 4 SWEPT BP selected.
Pressing the EXIT softkey displays the Hardware Diagnostic Mode menu.
Output signal routing for IPSI and CONTRA channels
by pressing the ROUTING softkey. The Routing menu
will be displayed.
The desired channel routing can be selected by press-
ing the corresponding softkey, as shown below:
I – I:
IPSI channel to IPSI transducer
I – C:
IPSI channel to CONTRA transducer
C – I:
CONTRA channel to IPSI transducer
C – C: CONTRA channel to CONTRA transducer
Grason-Stadler
ROUTING
4 - 14
Summary of Contents for TYMPSTAR 1
Page 1: ...Part Number 2000 0110RevB...
Page 20: ...Chapter 1 1 8 Grason Stadler...
Page 72: ...Chapter 3 Grason Stadler 3 32...
Page 168: ...Chapter 5 Grason Stadler 5 30...
Page 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Page 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Page 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Page 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Page 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Page 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Page 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Page 178: ...Chapter 6 Grason Stadler 6 10...
Page 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Page 180: ...Chapter 6 Grason Stadler 6 12...