Calibration
IPSI SOURCE
The signal source for the IPSI channel can be selected by press-
ing the IPSI SOURCE softkey. The possible IPSI sources are
shown in the center of the LCD in the IPSI list.
The desired source can be selected by scrolling through the list using the UP and
DOWN NO CAPS softkeys. This example shows the 1K Hz source selected.
Pressing the EXIT softkey displays the Hardware Diagnostic Mode menu at the
bottom of the screen.
IPSI ON/OFF
The IPSI output channel is toggled ON or OFF by pressing the
IPSI ON/OFF softkey.
CONTRA SOURCE
The signal source for the CONTRA channel can be selected
by pressing the CONTRA SOURCE softkey. The possible
CONTRA sources are shown in the center of the LCD in the
CONTRA list.
The desired source can be selected by scrolling through the list using the UP and
DOWN No CAPS softkeys. This example shows the 1K Hz source selected.
GSI TympStar Version 1 and Version 2 Service Manual
4 - 11
Summary of Contents for TYMPSTAR 1
Page 1: ...Part Number 2000 0110RevB...
Page 20: ...Chapter 1 1 8 Grason Stadler...
Page 72: ...Chapter 3 Grason Stadler 3 32...
Page 168: ...Chapter 5 Grason Stadler 5 30...
Page 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Page 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Page 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Page 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Page 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Page 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Page 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Page 178: ...Chapter 6 Grason Stadler 6 10...
Page 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Page 180: ...Chapter 6 Grason Stadler 6 12...