DFT Limit Too Big
While performing 2.0cc compliance calibration, the signal
processor ensures that the measured compliance at a nominal
gain setting allows for enough span in DFT calculation.
Subcodes: Out-of-range nominal DFT base.
Likely Causes: Clogged probe; mic circuit; filter circuit;
wrong cavity size used; bad probe SPL calibration data.
Invalid Speaker Level Input to Mux
Signal processor checks to ensure that when the probe
speaker level is to be written to EEPROM during pseudo-
code execution, that the probe speaker mux is set to the
corresponding position.
Subcodes: Invalid position.
Likely Causes: Software error.
Probe VCA DAC Level Out of Range
Signal processor checks to ensure that when the probe
speaker level is to be written to EEPROM when commanded
by the Control Processor (during Probe SPL calibration), that
the value is within a the valid range of 0 to 255 DAC counts.
Subcodes: The invalid level.
Likely Causes: Clogged probe tube; probe tone oscillator;
probe tone VCA.
Grason-Stadler
Chapter 8
SP Error codes
0041
0042
0043
8 - 18
Summary of Contents for TYMPSTAR 1
Page 1: ...Part Number 2000 0110RevB...
Page 20: ...Chapter 1 1 8 Grason Stadler...
Page 72: ...Chapter 3 Grason Stadler 3 32...
Page 168: ...Chapter 5 Grason Stadler 5 30...
Page 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Page 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Page 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Page 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Page 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Page 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Page 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Page 178: ...Chapter 6 Grason Stadler 6 10...
Page 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Page 180: ...Chapter 6 Grason Stadler 6 12...