Calibration
Verify Cal
Step 1
Press the VERIFY CAL softkey.
The PRESSURE CALIBRATION CHECK screen will be displayed.
Step 2
Follow the instructions on the screen to verify the pressure range linearity and safety
valve operation.
If the safety valve overpressure values fall outside the
acceptable range, open the TympStar enclosure
WARNING
This adjustment is performed while the instrument
enclosure is open and power is applied. Do not
touch wiring, circuitry or electrical compo-
nents! Voltages are present within the in-
strument enclosure that can cause serious
personal injury or death!
GSI TympStar Version 1 and Version 2 Service Manual
4 - 47
Summary of Contents for TYMPSTAR 1
Page 1: ...Part Number 2000 0110RevB...
Page 20: ...Chapter 1 1 8 Grason Stadler...
Page 72: ...Chapter 3 Grason Stadler 3 32...
Page 168: ...Chapter 5 Grason Stadler 5 30...
Page 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Page 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Page 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Page 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Page 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Page 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Page 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Page 178: ...Chapter 6 Grason Stadler 6 10...
Page 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Page 180: ...Chapter 6 Grason Stadler 6 12...