Calibration
Feed the stiff end of the cleaning floss into the tube and pull the floss completely
through the tube. Discard the used floss. Repeat this process for each of the re-
maining probe tubes. Do not reuse floss.
Step 3
Reconnect the tygon tubing to the probe
tubes. The center tube has the larger di-
ameter.
Contra phone
Step 1
Unplug the Contra phone from the jack
on the probe box and make sure the
main transducer opening is clear.
Step 2
Unplug the Contra phone cable from
the transducer by pulling the cable
connector straight away from the
transducer body.
Step 3
Use isopropyl alcohol and a clean tis-
sue or Q-tip to clean the two jacks in
the Contra phone body, and the two
Contra phone cable plugs.
GSI TympStar Version 1 and Version 2 Service Manual
4 - 5
Summary of Contents for TYMPSTAR 1
Page 1: ...Part Number 2000 0110RevB...
Page 20: ...Chapter 1 1 8 Grason Stadler...
Page 72: ...Chapter 3 Grason Stadler 3 32...
Page 168: ...Chapter 5 Grason Stadler 5 30...
Page 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Page 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Page 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Page 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Page 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Page 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Page 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Page 178: ...Chapter 6 Grason Stadler 6 10...
Page 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Page 180: ...Chapter 6 Grason Stadler 6 12...