The TympStar is calibrated using a combination of front panel hardkeys and LCD
panel softkeys.
Hardkeys are used as follows:
Stimulus Up/Down arrows-
Scrolls up or down through available
stimulus tone frequencies.
Intensity Up/Down arrows-
Adjusts output level of stimulus tone up
or down in 0.5dB increments.
Present bar-
Toggles the stimulus tone on and off.
Data transfer key-
Stores currently displayed calibration
values of tone frequency and level in the
TympStar’s memory.
Return key (LCD panel)-
Returns to previous (higher) level in the
calibration menu hierarchy.
RETURN
hardkey
Grason-Stadler
Chapter 4
Calibration procedures
TympStar calibration
controls
Hardkeys
Softkey
functions
4 - 28
Summary of Contents for TYMPSTAR 1
Page 1: ...Part Number 2000 0110RevB...
Page 20: ...Chapter 1 1 8 Grason Stadler...
Page 72: ...Chapter 3 Grason Stadler 3 32...
Page 168: ...Chapter 5 Grason Stadler 5 30...
Page 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Page 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Page 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Page 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Page 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Page 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Page 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Page 178: ...Chapter 6 Grason Stadler 6 10...
Page 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Page 180: ...Chapter 6 Grason Stadler 6 12...