Specifications
Sensitivity ranges
The following admittance measurements give maximum range at standard probe
tone frequencies. Compliance “Y”, 226 Hz is measured in ml. All other units are in
mmhos.
(1 acoustic mmho = 10
-8
m
3)
/Pa.s)
Table 1: Tymp Mode (Y, B, G)
Frequency
Digital Read-Out
Graphical
Including Cursor
Display
226 Hz
-7.0 to +7.0
-1.0 to +7.0
678 Hz
-21 to + 21
-5.0 to +25
1000 Hz
-30 to +30
-5.0 to +30
Accuracy:
226Hz:
0.1 ml or 5%, whichever is greater.
Above 226Hz: (F/226) x 0.1 mmho or K% whichever is greater.
K FACTOR:
From 250 to 1500 Hz = 5%
Above 1500 Hz =10%
Table 2: Reflex Mode (Y, B, G)
Frequency
Digital Read-Out Graphical
Including Cursor Display
226 Hz
-7.0 to +7.0
-0.16 to +0.80
+0.16 to -0.80
678 Hz
-21 to + 21
-0.48 to +0.80
+0.48 to -0.80
1000 Hz
-30 to +30
-0.64 to +0.80
+0.64 to -0.80
Accuracy: 226 Hz is 0.02 ml or 5%,whichever is greater.
GSI TympStar Version 1 and Version 2 Service Manual
2 - 11
Summary of Contents for TYMPSTAR 1
Page 1: ...Part Number 2000 0110RevB...
Page 20: ...Chapter 1 1 8 Grason Stadler...
Page 72: ...Chapter 3 Grason Stadler 3 32...
Page 168: ...Chapter 5 Grason Stadler 5 30...
Page 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Page 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Page 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Page 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Page 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Page 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Page 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Page 178: ...Chapter 6 Grason Stadler 6 10...
Page 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Page 180: ...Chapter 6 Grason Stadler 6 12...