![GSi TYMPSTAR 1 Service Manual Download Page 96](http://html1.mh-extra.com/html/gsi/tympstar-1/tympstar-1_service-manual_2275118096.webp)
Pressing the RESUME softkey restarts the TympStar system and displays the Diag-
nostic Mode menu.
Grason-Stadler
Chapter 4
WATCHDOG TIMER
Pressing the WATCHDOG TIMER softkey displays the watchdog timer test in-
structions and test results.
Pressing the smaller time value softkey tests the system for premature watchdog
timer triggering. The software timer should complete its cycle without triggering the
watchdog timer error.
Pressing the larger time value softkey tests the system for watchdog timer failures.
The software timer cycle is long enough to trigger the watchdog timer, and should
trigger an error. When the error is triggered, an error message is displayed that
contains instructions. Follow the instructions on the LCD to PRINT control infor-
mation or RESUME operation.
4 - 24
Summary of Contents for TYMPSTAR 1
Page 1: ...Part Number 2000 0110RevB...
Page 20: ...Chapter 1 1 8 Grason Stadler...
Page 72: ...Chapter 3 Grason Stadler 3 32...
Page 168: ...Chapter 5 Grason Stadler 5 30...
Page 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Page 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Page 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Page 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Page 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Page 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Page 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Page 178: ...Chapter 6 Grason Stadler 6 10...
Page 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Page 180: ...Chapter 6 Grason Stadler 6 12...