Calibration
Step 4
Press the GO TO +400 softkey. The meter will display the c400 daPa
pressure. The manometer and LCD meter should agree within 10%.
Step 5
If the LCD meter is not within the required 10%, adjust the span potentiometer to
achieve an acceptable reading.
NOTE
The span potentiometer controls the -600 daPa and +400 daPa levels. Any
adjustment at one pressure affects the other. The adjustment of -600 and
+400 is sometimes an iterative process of adjusting one, then the other
until both ends of the range are acceptable.
Leak rate
Step 1
Press the LEAK RATE softkey. The LEAK RATE TEST screen will be displayed.
GSI TympStar Version 1 and Version 2 Service Manual
4 - 45
Summary of Contents for TYMPSTAR 1
Page 1: ...Part Number 2000 0110RevB...
Page 20: ...Chapter 1 1 8 Grason Stadler...
Page 72: ...Chapter 3 Grason Stadler 3 32...
Page 168: ...Chapter 5 Grason Stadler 5 30...
Page 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Page 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Page 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Page 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Page 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Page 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Page 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Page 178: ...Chapter 6 Grason Stadler 6 10...
Page 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Page 180: ...Chapter 6 Grason Stadler 6 12...