Calibration
Adjusting Cal HL and
The Cal HL level shown on the CONTRA SPL CAL and IPSI SPL CAL screens
Target SPL levels
can be adjusted in 5.0dB increments to accommodate background noise levels in
the calibration environment. As the Cal HL level is adjusted up or down, the TARG
SPL level automatically tracks to eliminate the need for separate transducer target
SPL adjustments.
80 dB HL changed to 90 dB HL
The TARG SPL level shown on the CONTRA SPL CAL and IPSI SPL CAL
screens can only be adjusted when a custom transducer has been specified by set-
ting switch #2 of the calibration options DIP switch to the ON position. The appro-
priate target SPL can be incremented up or down in 0.5dB increments for each
probe tone frequency.
NOTE
Target SPL values must be within 10dB of the GSI default values.
GSI TympStar Version 1 and Version 2 Service Manual
4 - 31
Summary of Contents for TYMPSTAR 1
Page 1: ...Part Number 2000 0110RevB...
Page 20: ...Chapter 1 1 8 Grason Stadler...
Page 72: ...Chapter 3 Grason Stadler 3 32...
Page 168: ...Chapter 5 Grason Stadler 5 30...
Page 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Page 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Page 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Page 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Page 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Page 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Page 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Page 178: ...Chapter 6 Grason Stadler 6 10...
Page 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Page 180: ...Chapter 6 Grason Stadler 6 12...