Troubleshooting
CP Error codes
3303
ABORT_HW_SEREE_WAIT_FOR_ACK
SEREE never acknowledged request for write
Subcodes: none
Likely Causes: As each byte is shifted out to the EE, the CP
software checks to ensure that the EE is ready to accept a
write request. The most likely cause for this would be a
faulty EE or a problem with the data/control lines that run
from the EE on the probe to the Digital board. Noise on
these lines would also be a problem.
3304
ABORT_HW_SEREE_NACK
EE NACKed the previous byte write.
Subcodes: none
Likely Causes: As each byte is shifted out to the EE, the CP
software checks the ACK line to ensure that writes are
proceeding correctly. The most likely cause for this would be
a faulty EE, or problems with the data/control lines that run
from the EE on the probe to the Digital board. Noise on
these lines would also be a problem.
3306
ABORT_HW_SEREE_PROBE_VER_INCOMPAT
Probe EE was incompatible version - could not upgrade.
Subcodes: current probe ver #
Likely Causes: An unknown probe version has been attached
to the TympStar. For example, a future release 3 probe
attached to a release 2 system. The release 2 system cannot
upgrade this format probe. A probe with bad data could
result in this failsafe, but only if it managed to get through the
probe-present check and hamming code to ensure that the
integrity of the EE was good.
3400
ABORT_KEY_NEWSK_SETOPMODE_SEREE
Fetch of Serial EE shadow took too long
Subcodes: none
Likely Causes: A maximum time of 30 seconds is put on the
probe EE shadow read. This code can indicate a large
number of retries with the SEREE read and can indicate a
problem with the EE or the data/control path back to the CP
(including noise that may be causing retried reads). Less
likely is a software problem that sees the shadow read task
failing and this would indicate an internal software failure.
GSI TympStar Version 1 and Version 2 Service Manual
8 - 15
Summary of Contents for TYMPSTAR 1
Page 1: ...Part Number 2000 0110RevB...
Page 20: ...Chapter 1 1 8 Grason Stadler...
Page 72: ...Chapter 3 Grason Stadler 3 32...
Page 168: ...Chapter 5 Grason Stadler 5 30...
Page 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Page 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Page 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Page 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Page 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Page 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Page 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Page 178: ...Chapter 6 Grason Stadler 6 10...
Page 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Page 180: ...Chapter 6 Grason Stadler 6 12...