![GSi TYMPSTAR 1 Service Manual Download Page 82](http://html1.mh-extra.com/html/gsi/tympstar-1/tympstar-1_service-manual_2275118082.webp)
Chapter 4
The gain for the desired circuit can be adjusted by pressing the corresponding softkey.
The selected circuit will be indicated in the left column of the DAC LEVELS portion
of the LCD screen with its level shown in the right column. The example below
shows the contra circuit selected with a level of 255.
NOTE
The IPSI/CONTRA/PROBE outputs must be turned ON using the IPSI ON/
OFF, CONTRA ON/OFF or PROBE ON/OFF softkey.
The minimum gain for each circuit corresponds to a DAC value of 0, the maximum
corresponds to 255.
The DAC Level is adjusted by pressing the UP and DOWN ARROW softkeys.
The adjustment increment can be set to step values of 1 or 10 by pressing the
STEP=1 or STEP=10 softkey.
Pressing EXIT returns the display to the DAC level menu, where other circuits can
be selected and adjusted in the same manner.
Pressing the RETURN hardkey displays the Hardware Diagnostic Mode menu at
the bottom of the screen.
Grason-Stadler
4 - 10
Summary of Contents for TYMPSTAR 1
Page 1: ...Part Number 2000 0110RevB...
Page 20: ...Chapter 1 1 8 Grason Stadler...
Page 72: ...Chapter 3 Grason Stadler 3 32...
Page 168: ...Chapter 5 Grason Stadler 5 30...
Page 170: ...Chapter 6 Case chassis Grason Stadler 6 2...
Page 171: ...System Level Parts GSI TympStar Version 1 and Version 2 Service Manual 6 3...
Page 172: ...Chapter 6 Overallassembly Grason Stadler 6 4...
Page 173: ...System Level Parts LCD assembly GSI TympStar Version 1 and Version 2 Service Manual 6 5...
Page 174: ...Chapter 6 Top case assembly Grason Stadler 6 6...
Page 176: ...Instrumentassembly 6 8 Chapter 6 Grason Stadler...
Page 177: ...System Level Parts Labels GSI TympStar Version 1 and Version 2 Service Manual 6 9...
Page 178: ...Chapter 6 Grason Stadler 6 10...
Page 179: ...System Level Parts PrinterAssembly GSI TympStar Version 1 and Version 2 Service Manual 6 11...
Page 180: ...Chapter 6 Grason Stadler 6 12...