Caution
Failure to follow the prescribed cleaning or disinfection procedures will void the probes warranty. Do not soak
or wipe the lens with any product not listed in the Voluson E-Series Basic User Manual and/or care card.
Doing so could result in irreparable damage to the probe.
Caution
Disinfect a defective probe before you return it. Be sure to tag the probe as being disinfected.
Caution
Follow the Care Card instructions supplied with each probe (inside the probe boxes) for disinfectants and
gels that are compatible with the surface material of the probes.
Note
For the latest list of compatible cleaning solutions and disinfectants refer to: http://www.gehealthcare.com/
usen/ultrasound/products/probe_care.html.
Please be aware of the sensitive
probe head. TAKE EXTREME
CARE!
NEVER place or store a probe on its
scan head!
When disinfecting a probe, ensure
that there is sufficient space between
the probe and the container bottom!
10.4.5.4 Disinfection by means of the RIC-Holder
Especially for Real-time 4D endocavity probes (RIC), it is necessary to take extreme care when transporting
the system with the probe attached, or during the disinfection process. Inadequate handling may lead to
dead elements, shocked head mechanics, etc.
The RIC-Holder (especially developed for RIC Real-time 4D endocavity probes) guarantees that the
sensitive probe head does not hit the bottom of the container during the disinfection procedure.
Note
Instructions are supplied with each RIC-Holder (KTZ225469).
10.5 Using a Phantom
The use of a Phantom is not required during Preventive Maintenance. Customer may use it as part of their
Quality Assurance Program tests.
Care and Maintenance
Voluson E-Series Service Manual
KTD106657 Revision 2
10-9
Summary of Contents for H48681XB
Page 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Page 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Page 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Page 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Page 365: ......
Page 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...