3.1.2.2 System Acclimation Time
After being transported, the Voluson E-Series system may be very cold or hot. It requires one hour for each
2.5°C increment if it's temperature is below 10°C or above 40°C.
Caution
Equipment damage possibility. Turning the system on without acclimation after arriving at site may cause the
system to be damaged.
Table 3-2 Acclimation Time
°C 60
55
50
45
40 35 30 25 20 15 10 5
0 -5 -10 -15 -20 -25 -30 -35 -40
°F 140 131 122 113 104 96 86 77 68 59 50 41 32 23 14
5
-4 -13 -22 -31 -40
hrs
8
6
4
2
0
0
0
0
0
0
0
2
4
6
8
10 12 14 16 18 20
3.1.2.3 Control Console Positioning
If weight is placed on the control console (UI) in it's extended position the system could tip over.
Caution
The system should not be moved with the control console (UI) extended. Move the control console to it's
centered and locked position for transport.
Caution
Monitor mounting mechanism may break if not properly supported (e.g., with packing foam) during
transportation.
3.1.2.4 Brake Pedal Operation
Caution
If the wheel brakes are engaged, release brake pedals (brakes on wheels under the foot rest) to disengage
the lock, for transportation.
Setup Instructions
Voluson E-Series Service Manual
KTD106657 Revision 2
3-3
Summary of Contents for H48681XB
Page 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Page 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Page 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Page 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Page 365: ......
Page 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...