12. Updating will take some time ....
First the image is saved to R:\ (
Saving NTSF to image ....
); then the image is restored from R:\
(
Restoring NTFS from image ....
).
Figure 8-16 Please wait ...
13. Please wait until all processes are finished (100 percent completed).
14. The system is rebooting into windows (Boot screen - Voluson).
Note
An automated process was developed to install the required software parts, perform check disk, remap drive
letters and match settings. During this process the system might reboot several times!
Figure 8-17 automatic processes are running
15. Please wait until all processes are finished.
Replacement Procedures
Voluson E-Series Service Manual
KTD106657 Revision 2
8-11