8.14 Replacement of the Caps for TGC Sliders and/or Rotation Digipots
Note
The FRU part depends on Console (RTU). The console version (Rafi or Whanam) is shown in the System
).
Manpower
One person, 10 minutes
Tools
none (poss. small-sized slotted screwdriver or tweezers)
8.14.1 Caps for TGC Sliders and/or Rotation Digipots - Replacement Procedure
1.
Remove the caps for Slider-potentiometer TGC and/or Rotation Digipots.
Note
Do not loose integrated metal spring in each Rotation Digipot cap.
Figure 8-28 remove caps
2.
Mount the caps for Slider-potentiometer TGC and/or Rotation Digipots.
Replacement Procedures
Voluson E-Series Service Manual
KTD106657 Revision 2
8-23