2.2.3 Desirable Features
•
door is at least 92 cm (3 ft.) wide
•
circuit breaker for dedicated power outlet is easily accessible
•
sink with hot and cold water
•
receptacle for bio–hazardous waste, like used probe sheaths
•
emergency oxygen supply
•
storage for linens and equipment
•
nearby waiting room, lavatory, and dressing room
•
dual level lighting (bright and dim)
•
lockable cabinet for software and manuals
2.2.4 Network Setup Requirements
2.2.4.1 Stand-alone System (without Network Connection)
None
2.2.4.2 System connected to Hospital's Network
Supported networks:
•
Ethernet network connection
•
Wireless LAN (option)
2.2.4.3 Purpose of the DICOM Network Function
DICOM
®
1
(Digital Imaging and Communications in Medicine) services provide the operator with clinically
useful features for moving images and patient information over a hospital network. Examples of DICOM
services include the transfer of images to workstations for viewing or transferring images to remote printers.
As an added benefit, transferring images in this manner frees up the on-board monitor and peripherals,
enabling viewing to be done while scanning continues. With DICOM, images can be archived, stored, and
retrieved faster, easier, and at a lower cost.
2.2.4.4 DICOM Option Pre-installation Requirements
To configure the Voluson E-Series ultrasound system to work with other network connections, the network
administrator must provide some necessary information.
Use the
to record required information that must include:
•
Voluson E-Series Details: DICOM network details for the Voluson E-Series system, including the
host name, local port, IP address, AE title and net mask.
•
Routing Information:
IP addresses for default gateway and other routers in use at site.
•
DICOM Application
Information:
Details of DICOM devices in use at the site, including the DICOM host
name, AE title, DICOM port number and IP addresses.
Note
For further details refer to the Voluson E-Series Basic User Manual.
1 DICOM is the registered trademark of the National Electrical Manufacturers Association for its standards publications relating to digital communications
of medical information.
Site Preparation
2-8
Voluson E-Series Service Manual
KTD106657 Revision 2
Summary of Contents for H48681XB
Page 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Page 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Page 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Page 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Page 365: ......
Page 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...