6.3.2 Height Adjustment (Elevation)
Height adjustment is done with the Lift key on the control console. As long as the Lift key is pressed, the control
console can be lifted / lowered by means of the
up or down button displayed on the Touch Panel (soft keys for safety
reason).
Caution
Make sure that nothing would be jammed while moving!
6.3.2.1 Moving down (lower) the Console - without booting up the System
If it is impossible to boot up the system, the user interface can be moved downwards by pressing 3 keys on
the control console.
1.
Connect the main power cable to the back of the system.
2.
Connect the main power cable to a hospital grade power outlet with the proper rated voltage.
3.
) on the control console simultaneously to move it downwards.
Figure 6-4 keys to lower/lift the control console
Note
Shipping the Voluson E-Series ultrasound system in its original packaging is only possible when control
console is centered and locked in position, the system is lowered to its minimum height and the monitor is
6.3.2.2 Moving up (lift) the Console - without booting up the System
Note
Whanam console only: If it is impossible to boot up the system, the user interface can be moved upwards
(lifted) by pressing 4 keys on the control console.
1.
Connect the main power cable to the back of the system.
2.
Connect the main power cable to a hospital grade power outlet with the proper rated voltage.
3.
Press 3 keys + BF (see
) on the control console simultaneously to move it upwards.
Service Adjustments
6-6
Voluson E-Series Service Manual
KTD106657 Revision 2
Summary of Contents for H48681XB
Page 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Page 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Page 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Page 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Page 365: ......
Page 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...