4.4.4 Load Full System Configuration (Full Backup)
Caution
It is recommended to backup data before an upgrade; see
The “Full Backup” loading procedure replaces (overwrites) ALL existing data (except Application Settings
adapted for the used system software version) on the local hard drive of the Voluson E-Series system!
Note
It is neither required nor advisable to reload a previously stored “Full Backup” after a software upgrade that
was performed by means of the FMI from DVD button!
Note
There are circumstances where it is not possible to load (restore) all the data. The following rules specify
these restrictions:
1.
Generally, only restoring data from an older to a newer software version is possible. Loading a backup
into a system that has a lower software version than the system the backup was created on is
prohibited.
2.
Options can only be restored on the same Voluson E-Series system within the same major software
version.
3.
When loading a backup into a system with a software version that has a higher major number (13.x.x -
> 15.x.x), the following items will not be restored:
-
Options
-
State of the Service Platform
4.
The user is only allowed to restore data to a different system if and only if the software version on this
system is the same as in the backup.
5.
The user is not allowed to restore the following items to a different system:
-
Windows Network Settings
-
Options
-
DICOM AE Title and DICOM Station Name
-
State of the Service Platform
Operation
1.
Press the Utilities key on the control console.
2.
In the “Utilities” menu touch the
System Setup
button to invoke the setup desktop on the screen.
3.
On the left side of the screen select
Backup
and then click the
System Configuration
tab.
4.
Click the
Load
button of the "Full System Configuration" group.
5.
Choose the destination (1).
Figure 4-28 System Configuration Load
Functional Checks
4-36
Voluson E-Series Service Manual
KTD106657 Revision 2
Summary of Contents for H48681XB
Page 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Page 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Page 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Page 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Page 365: ......
Page 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...