Follow the test conditions described for respective test points shown in the table below.
Table 10-14 chassis leakage current test condition
Test
Condition
Test Point (image)
1
Potential equilibrium connector (rear of system, on power supply
RSP)
2
Monitor housing (mounting screw for LCD Monitor housing, rear of
monitor)
3
Probe connector (probe mounting receptacle on front of system)
10.6.5.3 Data Sheet for Enclosure/Chassis Leakage Current
shows a typical format for recording the enclosure/chassis leakage current. Measurements
should be recorded from multiple locations for each set of test conditions.
The actual location of the test probe may vary by system.
Record all data and keep the record of the results with other hard copies of maintenance data.
Table 10-15 Typical Data Format for recording Enclosure/Chassis Leakage Current
Unit under Test:
Date of Test:
Test Conditions
Measurement/Test Point Location
System Power
Grounding/PE
Potential equilibrium
connector
Monitor housing
Probe connector
off
closed
off
open
on
closed
on
open
Care and Maintenance
10-16
Voluson E-Series Service Manual
KTD106657 Revision 2
Summary of Contents for H48681XB
Page 11: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 3 ...
Page 12: ...Introduction 1 4 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 13: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 5 ...
Page 14: ...Introduction 1 6 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 15: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 7 ...
Page 16: ...Introduction 1 8 Voluson E Series Service Manual KTD106657 Revision 2 ...
Page 17: ...Introduction Voluson E Series Service Manual KTD106657 Revision 2 1 9 ...
Page 365: ......
Page 366: ...GE Healthcare Austria GmbH Co OG Tiefenbach 15 4871 Zipf Austria www gehealthcare com ...