GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
8-46
Section 8-3 - Internal Component Replacement Procedures
7.) Re-install the battery as described in the
"Battery Installation Procedure" on page 8-5
.
8.) Return the bearing handle as described in the
"Bearing Handle Installation Procedure" on page 8-3
.
9.) Turn ON the system and boot-up.
10.)Perform a complete functional check of the system, as described in
.
11.)Proceed to perform the following functionality tests:
-
"SafeLock Cart - Grounding Continuity" on page 10-22
- only if the
Vivid q N system is currently mounted on the SafeLock cart
-
"SafeLock Cart - Chassis Current Leakage Test" on page 10-24
-
"Calibration Tests" on page 7-25
-