GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
7-112
Section 7-6 - Performing Back End Diagnostics on the System
7-6-1-2
ECG Test
NOTE:
For a description of this test, refer to the information in
.
NOTE:
When performing this test, no probe should be connected to the system and nothing should be
touching the probe connectors!
1.) Connect the ECG cable to the input on the ECG port on the right side of the Vivid q N scanner.
2.) Connect the ECG leads to yourself (as the simulated patient).
3.) In the
Diagnostic
Test
window, trackball to
ECG Test
in the list of
BEP
tests available
for selection
from the Diagnostic Test Tree (see
) and press
Set
. The ECG Test name is
displayed under
Execute;
the relevant system components to be tested are highlighted light blue in
the Schematic View (BEP, Peripherals, Keyboard), as shown below:
4.) Under the
Special Instructions
area, activate the
Loops
checkbox
and select the required number of
Loops
from the combo box.
5.) Adjacent to
On Fail,
select either
Stop
or
Continue
, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
6.) Trackball to the
Start
button and press
Set
.
The interactive ECG Device Test screen is displayed as shown in
Figure 7-77 ECG Test