GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
8-30
Section 8-3 - Internal Component Replacement Procedures
8-3-1-4
P&M (Probe and MUX) Board Removal Procedure
1) Remove the bearing handle, as described in the
"Bearing Handle Removal Procedure" on page 8-2
.
2.) Remove the Control Panel and Keyboard, as described in the
Removal Procedure" on page 8-10
3.) Loosen and remove the four screws that secure the P&M Board to the RFI Board, as shown in
.
4.) Loosen the two screws that fasten the P&M Board to the Bottom Assembly, as shown in
Figure 8-31 Removing the Screws that Secure the P&M Board to the RFI Board
Figure 8-32 Loosening the Screws from the P&M Board