Chapter 7 - Diagnostics/Troubleshooting
7-97
5.) At any stage, trackball to the
Pause
button and press
Set
, if required.
6.) To resume testing (from the point where you paused), trackball to the
Start
button and press
Set
.
When the Temperatures Test sequence is complete:
-
the
Diagnostic
Test
window displays the Finished message
-
the View Test Log window opens automatically, listing all tests performed and showing details
of which tests passed or failed, as shown in the example in
.
NOTE:
For this test, there is one report (Temperatures) which is listed in the Reports window on the
right. By default, this report is displayed in the Reports window below.
7.) Check that the test passed and that result values are within the pre-defined
Max
and
Min
limits, as
shown in
.
8) When done, trackball to the
Exit
button and press
Set
to close the View Report window.
Figure 7-67 View Test Log - Temperatures Test