Chapter 8 - Replacement Procedures
8-51
8-3-6
Fan Replacement Procedure
NOTE:
This section describes the replacement procedure for the Fan Assembly which is supplied as
part of the Fans and Speaker Kit - Part No. 2412826.
8-3-6-1
Tools
Use the appropriate flat and Phillips-type screw drivers, as indicated in the fan replacement procedure.
8-3-6-2
Time Required
15 minutes
8-3-6-3
Preparation
Shut down the Vivid q N ultrasound unit, as described in
8-3-6-4
Fan Removal Procedure
1) Remove the bearing handle, as described in the
"Bearing Handle Removal Procedure" on page 8-2
.
2.) Remove the Control Panel and Keyboard, as described in the
Removal Procedure" on page 8-10
3.) Unscrew and remove the two screws that secure the Fan Assembly to the Bottom Assembly as
shown in
Figure 8-59 Removing the Screws from the Fan Assembly
4.) Disconnect the speaker cable from the fan assembly, as shown in
Figure 8-60 Disconnecting the Speaker Cable