Chapter 8 - Replacement Procedures
8-21
8-2-5-5
Rear Cover & Latch Assembly Installation Procedure
1.) Return the Rear Cover and Latch Assembly to the system and place it in position on the Bottom
Assembly, carefully checking that the
left
and
right
hinges are properly aligned.
2.) Return and fasten the two screws on each of the
left
and
right
hinges (previously removed - refer
3.) Return the LCD Display; see the
"AUO LCD Display Installation Procedure" on page 8-57
.
4.) Return the LCD Display Frame as described in the
"LCD Display Frame Installation Procedure" on
5.) Return the Control Panel and Keyboard as described in the
Installation Procedure" on page 8-12
6.) Return the Bearing Handle as described in the
"Bearing Handle Installation Procedure" on page
.
7.) Reconnect the AC adapter and all cables to the Vivid q N portable ultrasound scanner.
8.) Proceed to perform the following functionality tests:
-
"SafeLock Cart - Grounding Continuity" on page 10-22
- only if the Vivid q N system is
currently mounted on the SafeLock cart
-
"SafeLock Cart - Chassis Current Leakage Test" on page 10-24
-