GE
D
IRECTION
FQ091019, R
EVISION
2
V
IVID Q
N S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-47
8-3-4
BIOS Battery Replacement
NOTE:
Since the BIOS battery is not available as a spare part (due to shelf life), it should be obtained
locally. The battery model is CR2032.
8-3-4-1
Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the BEP replacement procedures.
8-3-4-2
Time Required
15 minutes
8-3-4-3
Preparations
Shut down the Vivid q N ultrasound unit, as described in
8-3-4-4
BIOS Battery Removal Procedure
1) Remove the BEP, as described in the
"BEP Removal Procedure" on page 8-40
.
2.) Take out the BIOS battery located on the bottom of the BEP, as shown in
.
8-3-4-5
BIOS Battery Installation Procedure
NOTE:
Before proceeding, refer to the
"Preparation Procedure" on page 8-45
.
1) Fit a new BIOS battery in position on the BEP - refer to
2.) Re-install the BEP, as described in the
"BEP Installation Procedure" on page 8-44
.
3.) Turn ON the system and boot-up.
4.) Adjust to the correct date and time settings - see
"Date and Time Adjustments" on page 3-71
.
5.) Proceed to perform the following functionality test:
-
"Battery Parameters Test" on page 7-102
Figure 8-55 BEP - Location of BIOS Battery
BIOS Battery