52
The values for DAC [0 mA] and DAC [20 mA] are factory default values.
The correct values for DAC [0 mA] and DAC [20 mA] must be adapted by
the user.
4
When desired set mA values
for error and idle state with
Result Value / Special
Signals -> Current
›
Error
and
Idle
.
5
Close the dialog with
Set
.
Current Loop Settings
Values (Min, Max, Default and Example Values)
Result Value-Range
Min 0mA, Max 20 mA (Default: 4,0 mA
–
20 mA)
DAC-Range
Depending on Current Loop generation
Min 0 mA
–
Max 1024 mA (Example: 22 mA
–
981 mA)
Min 0 mA
–
Max 65535 mA (Example: 20 mA
–
54559 mA)
Error Value
Min 0 mA, Max 20 mA (Defalut: 3,2 mA)
Idle Value
Min 0 mA, Max 20 mA (any value except Error Value and Result
Value Range)