30
1
Inspect
spectrum
for
contamination.
Control
the
baseline. Start cleaning when
contaminated.
Contamination is indicated by
peaks or the disappearing of the
RIP.
2
Select:
System
›
Modes
›
Cleaning
›
Reduced Flows / High Flows
.
Reduced Flows
:
Applied
for
minor
contamination: Flows remain
unchanged
and
low
(150
mL/min. for IMS and 15 mL/min.
for GC column) in order to
minimize gas consumption from
socket
Driftgas
. Process takes
more time.
High Flows
:
Applied
for
stronger
contamination:
Higher
gas
consumption
from
socket
Driftgas
, faster cleaning process
(150mL/min IMS and 15 mL/min.
for GC-column).
Recommendation: High Flows