Measurement
–
Repetition Time
Time span between two measurements.
Calibration Button
When deactivated no calibration is performed
during the process.
Calibration
–
Start
Date and Time
Start time and date of the first calibration in
the process.
Note that a calibration has priority over
a measurement. When a calibration and
a measurement collide according to the
scheduling, the measurement is skipped.
Calibration
–
Repetition Time
Calibration interval.
Edit…
Button
Opens the Substance Calibration Dialog.
→ 17.5
Cleaning Button
When deactivated no Cleaning cycle is
performed during the process.
Cleaning
–
Start Date
and Time
Start time and date of the first cleaning in the
process.
Note that a cleaning has priority over a
measurement and over calibration. When
any of them, calibration or measurement,
collide with cleaning according to the
scheduling, the measurement and/or
calibration are skipped.
Cleaning
–
Repetition
Time
Cleaning Interval
Cleaning Duration
Duration of the cleaning time
Cooldown time
Estimated time required for the instrument to
cool-down after a Cleaning