128
Line
Description
1-6
Head area
1
The Head starts with the Result count, in this case 11.
2
A zero follows.
3,4
Then the date: 14950080066 seconds since 01.01.1970 00:00 .
5,6
Followed by more zeroes (future use).
7-12
First Result
7,8
Substance shortname:
Register 7: 0x53 = "S", 0x49 = "I"
Register 8: 0x4C = "L" , 0x43 = "C"
The concatenated shortname is "SILC".
9,10
Concentration value:
First integer 81
Then divisor 100
So the real concentration value is 0.81 with the same unit as
indicated on the device.
11,12
Status:
Here is 0 so everything went as expected.
13-18
Second Result
Structured like the first
After the last result the remaining Registers will contain zeroes
Summary of Contents for GC-IMS-SILOX
Page 1: ...G A S Gesellschaft f r analytische Sensorsysteme mbH GC IMS SILOX USER MANUAL...
Page 66: ...65 7 Close the dialog with Close The import process is completed...
Page 103: ...102 17 10 2 System Page Plan Tab Figure 22 The System Page Plan...
Page 128: ...127 Exemplary Readout on Modbus...
Page 136: ...135 Figure 31 Results Window...