138
Addendum for GAS Internal use
LABMODE – Short introduction
By Entering the Lab Mode the GC-IMS-SILOX Firmware Enables the full
Development Options which usually are hidden from the normal User to make
the Graphical User Interface more user-friendly and avoid operational errors.
Support @ G.A.S. Dortmund can provide a Temporary Password
The Labmode should be opened by a G.A.S. trained expert only.
LabMode is used :
-
To Edit/view Measurement Porgrams
-
to perform the calibration measurements and manage Substance
calibrations
-
to modify system settings etc
It is very important to respect the Definitions of the Application-Modes. To run
the Silox1 Application Mode it needs a Program
SILOX1_MEA
and a Program
SILOX1_CAL
. Also it is necessary to create the
SILOX1
related Substance
definitions and prepare the calibration models in the required Calibration.gsd
file. If any of the required Programs or definitions are missing, the application
mode becomes invalid and
no
measurement is possible. To warn the user the
instrument will switch in a
DUMMY
mode. The following Screenshots shown to
give an overview.