Error Registers -
High Voltage Errors
Register
Data Type
Description
510
32 Bit int (combined with
Register 511)
Bit 0-15
0 : ERROR_I2C_CON
6 : ERROR_SOFTWARE_WATCHDOG
7 : ERROR_MISSING_PARAMETER
9 : ERROR_WATCHDOG_RESET
10 : ERROR_INVALID_HARDWARECONFIGURATION
11 : ERROR_EEPROM_CORRUPTED
12 : ERROR_INTERNAL_ERROR
13 : ERROR_OVERVOLTAGE
14 : ERROR_OVERLOAD
15 : ERROR_BROCKEN_WIRE
511
32 Bit int (combined with
Register 510)
Bit 16-31
Not used
EPC Errors
Register
Data Type
Description
522
32 Bit int (combined with
Register 523)
Bit 0-15
0 : ERROR_CONNECTION
1 : ERROR_PRESSURE_LOW
2 : ERROR_PRESSURE_HIGH
3 : ERROR_FLOW1_L
4 : ERROR_FLOW1_H
5 : ERROR_FLOW2_L
6 : ERROR_FLOW2_H
523
32 Bit int (combined with
Register 522)
Bit 16-31
Not used