19
Figure 2: Flow-chart
The drift gas flow to the
IMS
sensor is controlled by the electronic pressure
control unit
EPC1
. The carrier gas flow to the column is controlled by
EPC2
.
Carrier gas and drift gas leave the device through the
Gas Out
outlet at the rear
side of the device, which should be connected to an appropriate waste gas
ventilation system.
IMS
sensor,
GC
column and
6-Port-Valve
with sample loop
are heated (
T1
,
T2
and
T3
respectively). The transfer line between
GC
and
IMS
sensor is heated by
T5
.