120
Temperature Control Accuracy: ± 0.1 °C
Data storage
Internal storage volume
Data transfer via LAN connection via SMB, SFTP
or TFTP (G.A.S. variant of TFTP)
Controls
6.4” TFT
Touchscreen
Pushable rotary knob
Gas chromatographic column
Gaschromatographic capillary column
Type: MXT-5
Film Thickness: 1.0 µm
Column Length: 30 m
ID: 0.32 mm
Column Oven
Operating Temperature: 35
–
80 °C
Temperature Display Accuracy: ± 1°C
Temperature Control Accuracy: ± 0.1 °C
Flow Control EPC1
Drift Gas Flow
Type: Differential pressure control
Input Pressure 3.0 bar (300 kPa)
Output Pressure Stability 0.01% typical
Output Pressure Linearity 0.05% typical
Operation flow rates: 0
–
500 mL/min
Flow Control EPC2
Carrier Gas Flow
Type: Differential pressure control
Input pressure 3.0 bar (300 kPa)
Output pressure stability 0.01% typical
Output pressure linearity 0.05% typical
Operation flow rates: 0
–
150 mL/min
Cleaning mode
IMS, column and sampling system are heated up
to > 80°C (~ 100 °C).