17
When the calculated concentration value exceeds the respective maximum
value
MaxVal
it is displayed as
“
›
MaxVal”
.
When the calculated concentration value is below the respective minimum value
it is displayed as
“
n.d.
”
standing for
“not detected”
.
The concentration value of compound substances like Total Siloxanes
is
derived from base substances (L2, L3
, …
). When one or more base
substance concentrations exceed their concentration ranges the value
for the compound substance is derived from the limits.
When for all base substances (L2, L3
, …
)
a “n.d.” is calculated the
result for the compound substances is “n.d.” as well.
of silicon ’Total Si’ and silica ’Total SiO2’ in GC-IMS-
Summary of Contents for GC-IMS-SILOX
Page 1: ...G A S Gesellschaft f r analytische Sensorsysteme mbH GC IMS SILOX USER MANUAL...
Page 66: ...65 7 Close the dialog with Close The import process is completed...
Page 103: ...102 17 10 2 System Page Plan Tab Figure 22 The System Page Plan...
Page 128: ...127 Exemplary Readout on Modbus...
Page 136: ...135 Figure 31 Results Window...