123
19.6
Modbus TCP Specification
General
Connect the Device using MODBUS TCP on Port 502
Resulttransfer via “ReadHoldingRegister” (FC03)
Register 1
–
125: Current value
Register 127 251: Last value
One Result consists of 125 registers
A Result consists of the Head Area followed by the individual results
Description of Head area
Register
Data Type
Value
1
16 Bit unsigned int
Number of individual results
2
16 Bit unsigned int
Measurement state bitfield
3
32 Bit int (combined
with Register 4)
Date and time of measurement encoded
in seconds since 1st January 1970. This
register contains the least significant
two bytes.
4
32 Bit int (combined
with Register 3)
Date and time of measurement encoded
in seconds since 1st January 1970. This
register contains the most significant
two bytes.
5
Reserved for future use
6
Reserved for future use