125
Meaning of status bits for individual Results
Value
Description
0
Everything OK
1
Substance concentration is above the upper calibration limit of
the device (at least one part of accumulated/calculated
substance)
2
Substance concentration is below the lower calibration limit of
the device (not detected)
4
The measurement is invald
8
No calibration for this substance present on this device
>8
Reserved for future use!
Note:
In the rare case that the user switches the Application mode
(Silox1 to Silox2 or vice versa)
–
The Result Table shown on the
Result Screen may not match with the Result transferred on
Modbus TCP