51
1
Open dialog:
System
›
Settings
›
Current
Loop
>
Settings…
.
2
With
Result Value / Special
Signals -> Current
›
Min
(
1
)
and
Max
(
2
) set the mapping
of the concentration range
(
3
) to the range 0 mA
–
20
mA.
Here:
0.1 mg/m
3
→
4.0 mA
7.3 mg/m
3
→
20.0 mA
3
With
Result Value / Special
Signals -> Current
›
DAC [0
mA]
and
DAC [20 mA]
calibrate to 0-20 mA by
measuring the output Value
in mA on an external display
unit. The values in the
Screenshot
are
preset
factory default values and
have to be finetuned by the
user.
Summary of Contents for GC-IMS-SILOX
Page 1: ...G A S Gesellschaft f r analytische Sensorsysteme mbH GC IMS SILOX USER MANUAL...
Page 66: ...65 7 Close the dialog with Close The import process is completed...
Page 103: ...102 17 10 2 System Page Plan Tab Figure 22 The System Page Plan...
Page 128: ...127 Exemplary Readout on Modbus...
Page 136: ...135 Figure 31 Results Window...