Workflow: Current Loop
Setup
12.1
Introduction
The device transmits the measured concentration of the main substance as
electrical current to the 0-20 mA-current loop interface.
This is done only in the interval mode. See
Application Mode
Output Substance
SILOX1
Total Si
SILOX2
Total Si
For further information on measurement concentration ranges see
The concentration range of the output substance is mapped to a sub-interval
within 0-20 mA.
Outside this sub-interval a mA-error value and a mA-idle value can be defined.
The error value is set when a measurement has failed to produce a valid
concentration.
The idle value is set when no valid concentration is available yet or when the
interval mode is not active.