26
6
Connect
nitrogen
(Purity 5.0 or better)
source to input of
supplied
molecular
sieve.
Recommended pressure at socket Operation Gas is 3 bar - 4 bar (300
kPa
–
400 kPa). 6 bar (600 kPa) must not be exceeded.
7
Connect
molecular
sieve outlet to gas
socket
Operating Gas
.
8
Separately
connect
sockets
Sample Out
and
Gas Out
to a
Non-
sucking exhaust
or to
out-of-doors
.
9
Switch on device.
Minimum bypass flow at sockets Calibration In and Sample In: 200
ml/min.
Max 6
N
2
Molecular
Sieve
Pressure
Reducer