-40-
6.2 Sensor Connections
6.2.1 Moniturb F (MTF)Turbidity: Dual Sensors
MTF DUAL sensor wire connections
The table below shows
forward scatter
sensor 1 and 2 detector connections to
the Messenger.
MONITURB F Sensor 1 and 2 (Forward Scatter) Flow cell(s)
Terminal
#
Wire
Color
Cable
ID
Signal
1
Brown
FS (12
°
) sensor 1
(–) Direct beam
2
White
FS (12
°
) sensor 1
(+) Direct Beam
3
Black
FS (12
°
) sensor 1
Shield (ground)
4
Yellow
FS (12
°
) sensor 1
(–) 12 Deg. Sensor
5
Green
FS (12
°
) sensor 1
(+) 12 Deg. Sensor
6
-
-
-
7
Brown
FS (12
°
) sensor 2
(–) Direct beam
8
White
FS (12
°
) sensor 2
(+) Direct Beam
9
Black
FS (12
°
) sensor 2
Shield (ground)
10
Yellow
FS (12
°
) sensor 2
(–) 12 Deg. Sensor
11
Green
FS (12
°
) sensor 2
(+) 12 Deg. Sensor
The Messenger PC uses an “EXTERNAL” or supplemental (Condor) power
supply to drive the lamps of both MoniTurb F sensors.
The following connections are made to the Power Supply Control Board (located
inside the Messenger PC enclosure on the lower right corner of the back panel).