-63-
6.6.4.1
Set LCD Display
This option is not applicable to the Messenger PC.
6.6.4.2
Set Analog Outputs
The system has FOUR independent and assignable Analog 0/4 -20 mA loops,
which can be connected to External Devices. Each output will operate from a
user-defined RESULT equation. The sub-menus allow you to set all the
necessary parameter values for Low and High limits and the required mode for
the analog output.
Over range:
Analog Outputs can flag abnormal conditions or upset conditions that can be
caused by Malfunction of Optics or by some material coating over the windows.
We can assign a range of mV for normal operation. If the range exceeds, a fault
is registered and an appropriate error message (OPTICS FAULT) will be
displayed on the LCD and Measurements Screen.
How to configure Analog Outputs?
To configure the Analog Output, Choose the
ANALOG OUT
(Out 1-4).
Choose
Output
Toggle Button (Not Assigned / Result1 / Result2 / Result3
/ Result4) to specify the user defined RESULT equation needs to be
assigned to ANALOG OUTPUT.
Choose
Mode
Toggle Button to Specify the Mode (0/4-20 mA) Loops.
Choose
Low
,
High
text boxes to specify the Low and High mV.
Choose
Overrange
toggle button to specify the channel (CH1...4) Assign
the range of mV for normal operation by choosing the
Low
mV and
High
mV.
Note:
In case any of the Analog Output is not used, make sure that
OUTPUT
,
OVERRANGE
parameters are
"NOT ASSIGNED".
Summary of Contents for MONITEK MESSENGER
Page 1: ...MESSENGER PC TRANSMITTER User s manual Release 1 2 September 15 2015...
Page 11: ...9 Installation dimensions for Messenger PC transmitter in fiberglass enclosure...
Page 12: ...10 Installation dimensions for Messenger PC transmitter in stainless steel enclosure...
Page 13: ...11 1 3 2 Transmitter and Sensor Wiring System Layout and Interconnections details...
Page 14: ...12...
Page 15: ...13...