-56-
6.6.2.1
Process Calibration
Build Calibration
View/Edit Calibration
Fit Data
Copy Fact Table
Build Calibration
For each Sensor, up to 8 different samples can be used during the calibration
process.
Select the Sensor number by choosing the
Sensor
Toggle Button.
Select the Sample number by choosing the
Sample
Toggle Button.
Enter the
User Value
(Ex: - Known concentration value of a Standard) in the text
box provided. This value can also be changed using
View/Edit Calibration
menu.
Choose
Read New Data
to capture the data.
Choose
Store New Value
to store the data.
Repeat steps 2 to 5 for each sample.
View/Edit Calibration
The selected Sensor number appears on the top. The data present in the
selected process calibration table can be viewed/edited using the following
option.
To edit
user value
or
data,
Tap/Click on the cell.
Choose
Graph
to view the curve fit based on the
Fit Type
selected.
Choose
Edit Sensor Equations
to alter the Sensor Equations.
Summary of Contents for MONITEK MESSENGER
Page 1: ...MESSENGER PC TRANSMITTER User s manual Release 1 2 September 15 2015...
Page 11: ...9 Installation dimensions for Messenger PC transmitter in fiberglass enclosure...
Page 12: ...10 Installation dimensions for Messenger PC transmitter in stainless steel enclosure...
Page 13: ...11 1 3 2 Transmitter and Sensor Wiring System Layout and Interconnections details...
Page 14: ...12...
Page 15: ...13...